top of page

Specifications

  • Infusion rate

0.1 to 1,200 ml/h.

  • Volume to be infused (VTBI)

1 to 9,999 ml.

  • Volume infused (VI)

0.1 to 9,999 ml.

  • Keep vein open (KVO) infusion rate

1 to 20 ml/h or set infusion rate if lower.

  • Bolus during infusion

Bolus rate 10 to 1,200 ml/h (steps 10 ml/h)
Bolus volume 0.1 to 100 ml.

  • Free flow protection
  • Alaris Safety Clamp included in all Alaris GP sets.
  • Air-in-line detection
  • Triggered by single bubble or accumulation.
  • Occlusion pressure (user selectable)
  • 50 to 800 mmHg.
  • 9 levels of mmHg.
  • Alarms, warnings, prompts and advisories
  • Integrated amber/red beacon indicator, audible alarm and

display covering the following alarm conditions: AC power
failure and battery.

  • Air-in-line (various messages). Set misload.

Wrong set. Door close incomplete. Lever open. Set not fitted.
Upstream or downstream occlusion. No flow. Flow error.
Flow sensor disconnection. Near End of Infusion. VTBI done.
Attention. Rate lock. Guardrails™ alerts.

  • Dimensions and weight

148 mm (w) x 225 mm (h) x 148 mm (d).
2.5 kg including carry handle, internal battery and integrated
pole clamp.

  • LCD display

60 mm (w) x 60 mm (h).

  • Viewing angle of 35°, legible from 3 m.
  • Drug names up to 20 characters.
  • Mechanism

4 finger peristaltic displacement.

  • Volume accuracy

±5% (nominal).

  • Event log

Capacity 100,000 events, approximately 1 year of event log
storage in normal use.

  • Battery specifications

Integrated rechargeable NiMH; mean battery life 6 hours at
25.0 ml/h; 2.5 hours from discharge to 95% charge; battery
status indicator included.

  • Power requirements

100-230 VAC, 50-60 Hz, 60 VA (maximum).

  • Additional specifications

IEC/EN 60601-1, IEC/EN60601-2-24; IP33; RS232;

  • Alaris Gateway compatible; suitable for ground ambulance

use.

BD Alaris GP Plus Volumetric Pump

    Related Products