  1. The Alaris CC Plus syringe pump offers a complete IV infusion and blood transfusion solution allowing for timely, precise and well monitored treatments
  2. Alaris CC Plus syringe pump is particularly suited for use in both adult and paediatric critical care areas.
  3. Alaris CC Plus syringe pump is designed to increase patient safety and prevent harmful IV medication errors
  4. Infusion rate range 5 ml syringe 0.1 to 150 ml/h; 10 ml syringe 0.1  300 ml/h; 20 ml syringe 0.1 to 600 ml/h; 30 ml syringe 0.1 to 900 ml/h; 50 ml syringe 0.1 to 1200 ml/h; increments starting at 0.01 ml/h.
  5. Volume to be infused (VTBI) 0.1 ml to 1000 ml, 1 min to 24 hours
  6. Volume over time range 0.1 to 9,990 ml
  7. Bolus rate 5 ml syringe 10 to 150 ml/h; 10 ml syringe 10 to 300 ml/h; 20 ml syringe 10 to 600 ml/h; 30 ml syringe 10 to 900 ml/h; 50 ml syringe 10 to 1200 ml/h
  8. Purge specifications (syringe size specific) 100 to 500 ml/h; volume range 0.5 - 5.0 ml
  9. Keep vein open (KVO) infusion rate 0.1 to 2.5 ml/h
  10. Flow rate accuracy (ml/h mode) ±2% for rates ≥ 1.0 ml/h in accordance with IEC 60601-2-24*.
  11. Pumping pressure limits (user selectable) In-line pressure measurement with pressure disc; variable pressure; 1 to 1000 mmHg in 1 mmHg steps; auto pressure; 12 hour pressure trend graph
  12. Alarms, warnings, prompts and advisories Integrated amber/red beacon indicator, audible alarm and display covering the following alarm conditions: AC power failure and battery. Drive disengaged. Check syringe. Near end of infusion. VTBI done. Internal malfunction. Pressure disc out.Occlusion. Battery low. End of infusion. Attention (nurse call-back). Titration - start to confirm. Guardrails™ alerts.
  13. Dimensions and weight 335 mm (w) x 121 mm (h) x 200 mm (d). 2.4 kg including carry handle, internal battery and integrated pole clamp.
  14. LCD display 35 x 167 mm high resolution LCD display with LED backlight. Minimum viewing angle of 35°, legible from 3 m. Drug names up to 20 characters.
  15. Configurable options Audio volume (low, medium, high). Enable/disable syringe brands and sizes (5 ml to 60 ml). Quiet mode, rate lock, titration and standby mode.
  16. Battery specifications NiMH – rechargeable and replaceable. Mean battery life 6 hours at 5.0 ml/h. Recharge time 2.5 hours from discharge to 90% charge.
  17. Power requirements 115-230 VAC, 50-60 Hz, 10 VA (nominal).
  18. Event log Capacity 50,000 events. Approximately 1 year of event log storage in normal use.
  19. Additional specifications IP32 (IP33 with AC Power cable retainer kit); RS232; suitable for ground ambulance use (EN 1789); Alaris Gateway compatible; configurable through Alaris Editor; up to 30 profiles and 3,000 drug protocols (100 per profile).